PCR – Ireland Clam RLO DNA S/6/14 #19 from 20150130

After the last PCR continued to exhibit products in the no template controls (NTC) for most of the primer sets I was using, I ordered new primers. They arrived today so, I re-ran the PCR on the clam RLO DNA isolated 20150130 with the following new primers:

Master mix calcs are here: 20150223 – cPCR Universal Primers Apex Red MM

Cycling params were:
1 cycle of:

  • 95C – 10mins

40 cycles of:

  • 95C – 15s
  • 50C – 15s
  • 72C – 1min

Samples were run on 1.0% agarose, low TAE gel stained w/EtBr.

Results:

Ladder: Hyperladder I (Bioline).

Crazy; contamination still present in the NTCs. Primer stocks were steriley reconstituted with Low TE Buffer (IDT) and working stocks were created steriley, so I’m not really sure why this is continuing to happen. Possibly the polymerase is contaminated?  Will try again with previously unopened polymerase and see how that plays out.

No bands were excised since I can’t be certain that the bands present in the Clam DNA samples are from the actual sample and not from the apparent contamination.

One thought on “PCR – Ireland Clam RLO DNA S/6/14 #19 from 20150130

  1. Pingback: PCR – Ireland Clam RLO DNA S/6/14 #19 | Sam's Notebook